Product Category

Fully automatic Karl Fischer Coulomb moisture analyzer

  • Category :
    instrument
  • Browse number : ...
  • Release time : 2024-05-28
  • Detailed Description

Fully automatic Karl Fischer Coulomb moisture analyzer
Swiss Wantong's new generation integrated Karl Fischer Coulomb moisture analyzer

The system integration is very high, and the appearance design is simple and fashionable

Capable of Coulomb method for moisture determination and bromine index determination

Built in suction and drainage device to ensure user safety

Can automatically perceive sample addition and start titration automatically

USB storage anti-counterfeiting PDF experimental report, which can be sent through the network

Multiple dialogue languages including Chinese

917 Coulometer Elf Generation One piece Karl Fischer Coulomb Moisture Tester

The 917 Coulomete Elf Generation is an integrated Karl Fischer Coulomb moisture analyzer launched by Swiss company Wantong for daily analysis. There are two modes available: Karl Fischer Coulomb moisture determination and bromine index determination.

She is compact and integrates an operation panel, magnetic stirrer, titration tank, and suction and discharge pump into one. The overall structure is compact and flexible, saving your valuable laboratory space.

After adding optional accessories, fully automated testing can be achieved

Connecting the 885 Cassette sample heating processor makes it easy to achieve measurement automation

◆ Paperless office, PDF report

Can generate anti-counterfeiting PDF experimental reports and store them in USB memory or network computers, fashionable and environmentally friendly

◆ Personalized touch screen, quick access to experimental methods

14 shortcut icons can be created on the touch screen to directly call methods for experimentation

◆ With reagent monitoring function, ensuring great safety

The reagents are continuously monitored, and once the total number of injections or the total amount of injected water meets the custom quantity limit, the instrument will automatically prompt the user to replace the reagents.

◆ Tap the button to empty and add reagents

The device is equipped with a built-in suction and discharge pump system, which allows you to empty the waste liquid and add reagents without opening the titration cup. The operator can completely avoid contact with the reagents.

◆ KF icon displays real-time device operation status

The KF icon on the touch screen displays in real-time whether the device is in pre titration state or in sample preparation state

◆ Automatically sense the sample and start moisture measurement

No need to manually start the moisture measurement. After the sample is added, the device can automatically sense the sample and start the measurement automatically

Flexible upgrade, can be used in conjunction with 860 and 885 Cassette sample heating processors

Not all samples can be directly injected. It is recommended to use it in conjunction with a Karl Fischer furnace when the following conditions exist in the samples.

◆ The sample cannot be completely dissolved (plastic particles, polyester chips, artificial bones, etc.)

◆ Side reactions between samples and Karl Fischer reagents (lubricant additives, samples containing aldehydes and ketones, etc.)

◆ Slow release of sample moisture

The moisture in the sample can only be released at high temperatures

◆ The sample will contaminate the titration cell (lubricating grease, etc.)

After being combined with a Karl Fischer furnace, the moisture in the sample is heated and separated, and then transferred to the titration cell through a dry carrier gas, where moisture measurement is performed.

Characteristics of the Karl Fischer furnace:

◆ Independent sealed sample bottle, suitable for on-site direct sampling and sealing

◆ No sample transfer is required during measurement, resulting in more accurate results

◆ Head space heating method, small dead volume, and low cross contamination of samples

860 type single grade Cassegrain heating furnace with optional heating chamber and sample bottle volume

885 type fully automatic Cassette heating furnace, with 18 sample positions, can achieve fully automatic determination of sample blank and sample water content

Example of application of Coulomb Karl Fischer moisture determination:

Petroleum and Petrochemical

Industrial product category

Food category

Pharmaceutical category

cosmetic

Organic heat carrier

Moisture in methanol and ethanol

Edible fats and oils

Active pharmaceutical ingredients

Lipstick

Engine oil, transformer oil

Coking products

milk powder

Finished medicine

Skincare products

Aviation kerosene

Resin, epoxy resin

coffee

Freeze dried powder injection

Sunscreen

lubricating oil

Biodiesel

butter

Biological products

 
Engine oil

Organic solvents

     
Light oil

inorganic salt

     
Heavy oil

PE、PA、PP、PVC

     
 

 

Technical parameter list:

Moisture measurement range

Recommended range: 10 μ g~200mg, maximum measurement limit: 1000mg

measurement accuracy 

± 0.3% (H2O ≥ 1000 μ g)

Maximum electrolysis rate of the electrode

2241 μ g H2O/min

Generating electrode current

100, 200, 400mA and auto adjustable

Indicating electrode polarization current

AC power: 5, 10, 20, and 30 μ A

Direct current: -125...+125 μ A

Indicating electrode potential measurement resolution

0.1 mv

KFC Karl Fischer Coulomb method

have

BRC Coulomb Method for Determining Bromine Value Bromine Index

have

Built in diaphragm pump

Yes, it can automatically add reagents and empty the waste liquid in the measuring cup

Dosimats, Dosinos,  Mixer interface

Two MSB interfaces, which can be used to connect to the dispensing driver of Wantong to achieve fully automatic reagent quantitative addition and emptying

Karl Fischer heating furnace

860 type single grade Cassette sample heating processor (optional)

885 type (18 sample grades) Cassette sample heating processor (optional)

Balance, printer, PC, PC keyboard, barcode reader

Through 2 USB ports, USB/RS-232 converter (optional)

Can generate anti-counterfeiting PDF reports without using a computer

have

USB memory storage experimental method, results, and PDF report

have

Network port

Yes, it can be connected to the network through a network interface for data archiving, printed on a network printer, and directly connected to the LIMS system

touch screen

90 mm x 120 mm, high-resolution color screen, real-time curve display

Method and sample information storage, result storage, database

have

Dialogue language

English, French, German, Spanish, Chinese

Callable shortcut icons

14

Automatically start titration

have

Operator/Expert level mode switching

have

Drift correction function

Manual/Automatic/Off

Recalculation function of results

have

User defined calculation formula function

have

Extraction time setting function

have

url : https://en.haohuahuagong.cn/product/161.html